Endress+Hauser CA80PH Liquiline System Orthophosphate Analyzer


  • Part Number: Endress+Hauser
  • Category: Analyzer Systems
  • Manufacturer: Endress+Hauser (E & H)
  • Product Origin: Germany / EU Germany / EU
  • Issue Date: 2026-11-26 17:35:28
  • SKU: IS-108289
  • ✓Certified Refurbished & Surplus — Save 30-60% vs New Price
  • ✓Fully Tested & Calibrated — 100% Functional Test & Report Included
  • ✓3-12 Months Warranty — Guaranteed Functional Warranty
  • ✓Fast Global Express Shipping — 3-7 Days Delivery via DHL / FedEx / UPS
Inquire via WhatsApp
New Product
On Request
✓ Original Factory Packaging
✓ Ship by: 2-4 Weeks
Never Used W/O Package
On Request
✓ Open-Box Surplus Stock
✓ Protective Industrial Packaging
✓ Ship by: 3-7 Business Days
Refurbish Grade A
On Request
✓ Inspected & 100% Function Tested
✓ 90%+ Like New Appearance
✓ Certified Refurbished
✓ Ship by: 5-10 Business Days
Refurbish Grade B
On Request
✓ Used, Repaired & Function Tested
✓ Ship by: 4-10 Business Days

Taxonomy Path: Analyzer Systems > Liquid Analytics > Liquid Analytical Systems

CA80PH Liquiline System orthophosphate analyzer from Endress+Hauser
The
Endress+Hauser CA80PH
is a colorimetric wet-chemical analyzer for the almost continuous determination of orthophosphate in liquid media, with blue-method and yellow-method versions and up to two measuring channels. It is a member of the Liquiline System CA80 family: a cabinet analyzer with an integrated controller, a graphic display and optional digital Memosens sensor inputs. our engineering team, an independent supplier in Muscat, supplies the CA80PH in Oman and the GCC, and this page explains what the analyzer measures, how it is configured and what to confirm before ordering.
Typical uses named in the manufacturer documentation: Monitoring and optimization of the treatment efficiency of municipal and industrial wastewater treatment plants, monitoring and optimization of aeration basins, regulation of precipitant dosage and monitoring of cooling water circuits.
CA80PH technical specifications
Parameter
Specification
Manufacturer / part number
Endress+Hauser / CA80PH
Measured variable
PO4, PO4-P, P2O5 [mg/l, ppm]
Measuring range
CA80PH-**E1: 0 to 2.5 mg/l PO4-P (blue); -**E2: 0.05 to 10 mg/l (blue); -**E3: 0.5 to 20 mg/l (yellow); -**E4: 0.5 to 50 mg/l (yellow)
Dilution option
E2 with dilution factor 50 reaches 2.5 to 500 mg/l; E4 with factor 20 reaches 10 to 1,000 mg/l
Maximum measured error
2 % of end of measuring range; dilution factor 5 adds 2.0 % (blue) or 3.0 % (yellow)
Limit of detection
0.015 mg/l PO4-P (blue method only)
Repeatability
Blue: ±2 % of display value + 0.01 mg/l; yellow: ±2 % of display value + 0.05 mg/l
Measuring interval
Blue: continuous (approx. 11 min), adjustable above 11 min; yellow: approx. 8 min, adjustable above 10 min
Sample requirement
22 ml per measurement without dilution module
Reagent requirement
Blue: approx. 50 µl per reagent per measurement; yellow: approx. 115 µl per measurement
Standard requirement
About 230 ml a month at a 48 h calibration interval
Maintenan

Delivery: UPS, FedEx and DHL Express
International delivery available
Same day shipping for in-stock items on location
Will negotiate custom freight & clearance details
Combined Shipping Policy: We combine shipments to reduce costs for customers purchasing multiple items. After completing your selection, request a consolidated proforma invoice including shipping and handling.
Warranty: All products are covered by Global Valve Hub — Industrial Supplies & Engineering Solutions.
We record a functional test video before shipping.
✓ 3–12 Months Functional Warranty on New, Surplus & Refurbished Parts.
Returns: No-hassle return and replacement policy.
Quick Quotes: Receive Price & Availability Today! Email sales@globalvalvehub.com or WhatsApp +1 332 322 6825.
Payment Options: Wire Transfer (T/T) PayPal Visa MasterCard

Related Products

Endress+Hauser
Request Inquiry
More Info
Endress+Hauser
Request Inquiry
More Info
Endress+Hauser
Request Inquiry
More Info
Endress+Hauser
Request Inquiry
More Info
Endress+Hauser
Request Inquiry
More Info
Endress+Hauser
Request Inquiry
More Info
↑