Taxonomy Path: Analyzer Systems > Liquid Analytics > Liquid Analysis Sensors
Endress+Hauser CPS72E Memosens ORP sensor for fast-changing redox
Endress+Hauser CPS72E measures oxidation-reduction potential rather than pH. A platinum measuring element reports the redox state of the liquid against an Ag/AgCl reference, over a span of minus 1500 to 1500 mV. The datasheet frames it for process technology and monitoring where ORP values change quickly or where electrode poisons such as hydrogen sulphide are present.
The reference system is a 3 M gel KCl bridge free of AgCl and equipped with an ion trap, so the silver lead stays protected longer. As with other 2.0 sensors, a Memosens head digitises the signal, stores calibration and sensor load, and passes data inductively. our engineering team is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.
CPS72E technical specifications
Parameter
Value
Measured variable
ORP, temperature
ORP range
minus 1500 to 1500 mV
Process temperature
0 to 140 C (32 to 284 F)
Process pressure
0.8 to 14 bar absolute (11.6 to 203 psi)
Minimum conductivity
10 microsiemens per cm, atmospheric pressure, no flow
Reference
Ag/AgCl, gel KCl 3 M, AgCl-free, with ion trap
Measuring element
Platinum
Junction
Ceramic, zirconium dioxide
Temperature sensor
NTC 30K
Installed length
120, 225, 360, 425 mm
Weight
40, 60, 90, 100 g
Shaft diameter
11.9 mm (0.47 in)
Materials
Glass shaft, FKM O-ring, PPS-GF process coupling
Plug-in head
Memosens, 16 bar relative
Ingress protection
IP 68, 10 m, 25 C, 45 days, 1 M KCl
Storage
0 to 50 C
Ex marking
ATEX II 1G Ex ia IIC T3/T4/T6 Ga
How CPS72E differs from other Memosens sensors
Where CPS91E and CPS41E report pH through glass, CPS72E carries a platinum element and a range stated in millivolts. Its 140 C ceiling is the highest in this group, above 135 C for CPS41E and 110 C for CPS91E, while its 14 bar pressure rating equals CPS91E. Compared with CPS92D, an older Orbipore ORP electrode with open aperture, CPS72E uses a ceramic junction with ion trap instead; its
| Delivery: | UPS, FedEx and DHL Express International delivery available Same day shipping for in-stock items on location Will negotiate custom freight & clearance details |
|---|---|
| Combined Shipping Policy: | We combine shipments to reduce costs for customers purchasing multiple items. After completing your selection, request a consolidated proforma invoice including shipping and handling. |
| Warranty: | All products are covered by Global Valve Hub — Industrial Supplies & Engineering Solutions. We record a functional test video before shipping. ✓ 3–12 Months Functional Warranty on New, Surplus & Refurbished Parts. |
| Returns: | No-hassle return and replacement policy. |
| Quick Quotes: | Receive Price & Availability Today! Email sales@globalvalvehub.com or WhatsApp +1 332 322 6825. |
| Payment Options: | Wire Transfer (T/T) PayPal Visa MasterCard |