Fluke 1625-2 GEO Earth Ground Tester with AFC and R*


  • Part Number: 1625-2
  • Category: Portable Test & Measurement
  • Manufacturer: Fluke
  • Product Origin: Germany / EU Germany / EU
  • Issue Date: 2026-11-26 17:35:28
  • SKU: IS-108430
  • ✓Certified Refurbished & Surplus — Save 30-60% vs New Price
  • ✓Fully Tested & Calibrated — 100% Functional Test & Report Included
  • ✓3-12 Months Warranty — Guaranteed Functional Warranty
  • ✓Fast Global Express Shipping — 3-7 Days Delivery via DHL / FedEx / UPS
Inquire via WhatsApp
New Product
On Request
✓ Original Factory Packaging
✓ Ship by: 2-4 Weeks
Never Used W/O Package
On Request
✓ Open-Box Surplus Stock
✓ Protective Industrial Packaging
✓ Ship by: 3-7 Business Days
Refurbish Grade A
On Request
✓ Inspected & 100% Function Tested
✓ 90%+ Like New Appearance
✓ Certified Refurbished
✓ Ship by: 5-10 Business Days
Refurbish Grade B
On Request
✓ Used, Repaired & Function Tested
✓ Ship by: 4-10 Business Days

Taxonomy Path: Portable Test & Measurement > Electrical Testers & Multimeters

Fluke 1625-2 GEO earth ground tester with AFC and R* measurement
The
Fluke 1625-2
is a handheld earth ground tester that performs all four classic earth measurements: 3- and 4-pole fall of potential, 4-pole soil resistivity, selective testing with one clip-on clamp and stakeless testing with two clamps. It sits above the 1623-2, which covers the same measurement methods but without the Automatic Frequency Control and R* functions. our engineering team supplies the 1625-2 in Oman and the GCC as an independent supplier in Muscat, and this page sets out what the instrument does, what the data sheet specifies and what to include when you request a quotation.
What distinguishes the 1625-2 from the 1623-2
Both testers share the one-button rotary switch concept, the IP56 case, USB data storage of up to 1,500 records and the professional carrying case. The data sheet lists two extra features for the 1625-2. Automatic Frequency Control (AFC) identifies existing interference on the ground system and chooses a measurement frequency that minimises its effect, giving a more accurate earth value on noisy sites. R* measurement calculates earth impedance at 55 Hz, which better reflects the resistance that a fault-to-earth current would see. The 1625-2 also offers adjustable limits for quicker pass or fail testing. The 1625-2 reads on a 4-digit (2999 count) display, against a 1999-count display on the 1623-2.
Technical specifications of the 1625-2
Parameter
Specification
Manufacturer / part number
Fluke / 1625-2 GEO
Measuring functions
Earth resistance 3- and 4-pole, with or without clip-on transformer; interference voltage and frequency; 2-pole resistance (ac) and 2- and 4-pole (dc)
Earth resistance (RE) method
Current and voltage measurement with probes per IEC 61557-5
Open-circuit voltage
20 / 48 V ac
Short-circuit current
250 mA ac
Measuring frequency
94, 105, 111, 128 Hz (manual or automatic AFC); 55 Hz in R* function
Noise rejection
120 dB (16-2/3, 50, 60, 400 Hz)
Interference volta

Delivery: UPS, FedEx and DHL Express
International delivery available
Same day shipping for in-stock items on location
Will negotiate custom freight & clearance details
Combined Shipping Policy: We combine shipments to reduce costs for customers purchasing multiple items. After completing your selection, request a consolidated proforma invoice including shipping and handling.
Warranty: All products are covered by Global Valve Hub — Industrial Supplies & Engineering Solutions.
We record a functional test video before shipping.
✓ 3–12 Months Functional Warranty on New, Surplus & Refurbished Parts.
Returns: No-hassle return and replacement policy.
Quick Quotes: Receive Price & Availability Today! Email sales@globalvalvehub.com or WhatsApp +1 332 322 6825.
Payment Options: Wire Transfer (T/T) PayPal Visa MasterCard

Related Products

325
Request Inquiry
More Info
Ti300+
Request Inquiry
More Info
718
Request Inquiry
More Info
355
Request Inquiry
More Info
TiS20+
Request Inquiry
More Info
700G06
Request Inquiry
More Info
↑