Fluke 8508A Reference Multimeter


  • Part Number: 8508A
  • Category: Portable Test & Measurement
  • Manufacturer: Fluke
  • Product Origin: Germany / EU Germany / EU
  • Issue Date: 2026-11-26 17:35:17
  • SKU: IS-101375
  • ✓Certified Refurbished & Surplus — Save 30-60% vs New Price
  • ✓Fully Tested & Calibrated — 100% Functional Test & Report Included
  • ✓3-12 Months Warranty — Guaranteed Functional Warranty
  • ✓Fast Global Express Shipping — 3-7 Days Delivery via DHL / FedEx / UPS
Inquire via WhatsApp
New Product
On Request
✓ Original Factory Packaging
✓ Ship by: 2-4 Weeks
Never Used W/O Package
On Request
✓ Open-Box Surplus Stock
✓ Protective Industrial Packaging
✓ Ship by: 3-7 Business Days
Refurbish Grade A
On Request
✓ Inspected & 100% Function Tested
✓ 90%+ Like New Appearance
✓ Certified Refurbished
✓ Ship by: 5-10 Business Days
Refurbish Grade B
On Request
✓ Used, Repaired & Function Tested
✓ Ship by: 4-10 Business Days

Taxonomy Path: Portable Test & Measurement > Process, Pressure & Temperature Calibrators

Overview
Fluke 8508A
is an 8.5-digit reference multimeter built specifically for metrology laboratories that need a transfer standard, not just a bench meter. Where a typical digital multimeter targets 6 or 6.5-digit resolution for general troubleshooting, the
Fluke 8508A
resolves DC voltage to ±3 ppm of reading, DC current to ±12 ppm, and resistance to ±7.5 ppm using a true-ohms measurement technique that cancels thermal EMF errors — the level of stability required to verify the output of a multi-product calibrator or serve as a laboratory’s working standard between primary reference recalibrations.
With engineering support in Muscat, our engineering team supplies the
Fluke 8508A
to accredited calibration laboratories and national metrology-adjacent standards rooms across Oman and the GCC, with technical application guidance for labs building ISO/IEC 17025 traceability chains.
Operating Principle
The
Fluke 8508A
uses a high-linearity analog-to-digital converter combined with an ultra-stable internal voltage reference to resolve measurements to 8.5 digits, and its true-ohms resistance function alternates current polarity and subtracts consecutive readings to cancel thermal EMF offsets that would otherwise corrupt low-value resistance measurements at this resolution. Ratio and difference measurement modes let a metrologist directly compare an unknown standard against a known reference without manually computing the delta, reducing operator-introduced error during standards intercomparison work.
Because every function — DC/AC voltage to ±1050V, DC current to ±20A, resistance to 20 GΩ, and temperature from −200°C to 660°C via PRT — is derived from the same internal reference architecture, the 8508A’s own specifications become the uncertainty budget baseline for every measurement made downstream in a calibration lab. The practical consequence is that a lab’s entire chain of traceability, from working standards down to field test equipment, ultimately rests on how well this on

Delivery: UPS, FedEx and DHL Express
International delivery available
Same day shipping for in-stock items on location
Will negotiate custom freight & clearance details
Combined Shipping Policy: We combine shipments to reduce costs for customers purchasing multiple items. After completing your selection, request a consolidated proforma invoice including shipping and handling.
Warranty: All products are covered by Global Valve Hub — Industrial Supplies & Engineering Solutions.
We record a functional test video before shipping.
✓ 3–12 Months Functional Warranty on New, Surplus & Refurbished Parts.
Returns: No-hassle return and replacement policy.
Quick Quotes: Receive Price & Availability Today! Email sales@globalvalvehub.com or WhatsApp +1 332 322 6825.
Payment Options: Wire Transfer (T/T) PayPal Visa MasterCard

Related Products

325
Request Inquiry
More Info
Ti300+
Request Inquiry
More Info
718
Request Inquiry
More Info
355
Request Inquiry
More Info
TiS20+
Request Inquiry
More Info
700G06
Request Inquiry
More Info
↑