JUMO AQUIS 500 Ci Inductive Conductivity Transmitter


  • Part Number: AQUIS
  • Category: Analyzer Systems
  • Manufacturer: JUMO
  • Product Origin: Germany / EU Germany / EU
  • Issue Date: 2026-11-26 17:35:28
  • SKU: IS-106539
  • ✓Certified Refurbished & Surplus — Save 30-60% vs New Price
  • ✓Fully Tested & Calibrated — 100% Functional Test & Report Included
  • ✓3-12 Months Warranty — Guaranteed Functional Warranty
  • ✓Fast Global Express Shipping — 3-7 Days Delivery via DHL / FedEx / UPS
Inquire via WhatsApp
New Product
On Request
✓ Original Factory Packaging
✓ Ship by: 2-4 Weeks
Never Used W/O Package
On Request
✓ Open-Box Surplus Stock
✓ Protective Industrial Packaging
✓ Ship by: 3-7 Business Days
Refurbish Grade A
On Request
✓ Inspected & 100% Function Tested
✓ 90%+ Like New Appearance
✓ Certified Refurbished
✓ Ship by: 5-10 Business Days
Refurbish Grade B
On Request
✓ Used, Repaired & Function Tested
✓ Ship by: 4-10 Business Days

Taxonomy Path: Analyzer Systems > Liquid Analytics > Liquid Analytical Systems

Maintenance-light conductivity and concentration control for dirty media
The
JUMO AQUIS 500 Ci
(type 202566) is a wall- or panel-mounted transmitter and controller that reads inductive conductivity cells and the temperature of the process liquid. Because an inductive cell has no metal electrodes in contact with the liquid, the measurement tolerates oil, grease, gypsum and lime deposits better than a conductive two-electrode cell, which makes the instrument a practical choice for CIP return lines, cleaning-chemical dosing and process water that fouls quickly. The unit converts the compensated conductivity either into a conductivity value in µS/cm or mS/cm or, with direct selection, into a concentration for NaOH, HNO3, H2SO4 or HCl, and it can also use a customer table of up to 29 value pairs entered through the optional setup program. A graphic LC display with plain-text parameters, a calibration logbook holding the last five calibrations and a binary input for HOLD or key inhibit keep daily operation simple.
Technical specifications
Parameter
Specification
Basic type
202566, JUMO AQUIS 500 Ci, transmitter/controller for inductive conductivity, concentration and temperature
Conductivity ranges
0 to 9999 µS/cm; 0.000 to 9.999, 0.00 to 99.99 and 0.0 to 999.9 mS/cm; 0 to 2000 mS/cm
Accuracy
1.5 % of range (0 to 1.000 mS/cm); 1 % of range (1.01 to 500 mS/cm); 1.5 % of range (501 to 2000 mS/cm)
Concentration ranges
NaOH 0 to 12 and 20 to 50 wt%; HNO3 0 to 25 and 36 to 82 wt%; H2SO4 0 to 28, 36 to 85 and 92 to 99 wt%; HCl 0 to 18 and 22 to 44 wt%
Temperature input
Pt100 or Pt1000 (automatic detection), -50 to 250 °C; NTC/PTC up to 4 kΩ; accuracy 0.5 K to 100 °C
Temperature compensation
Linear 0 to 5.5 %/°C (0 to 100 °C), natural water to ISO 7888 (0 to 36 °C), reference temperature 15 to 30 °C (25 °C preset)
Cell constant adjustment
4 to 6 and 6 to 8 per cm; relative cell constant 80 to 120 %; installation factor 80 to 120 %
Analog outputs (max. 2)
0/4 to 20 mA (load up to

Delivery: UPS, FedEx and DHL Express
International delivery available
Same day shipping for in-stock items on location
Will negotiate custom freight & clearance details
Combined Shipping Policy: We combine shipments to reduce costs for customers purchasing multiple items. After completing your selection, request a consolidated proforma invoice including shipping and handling.
Warranty: All products are covered by Global Valve Hub — Industrial Supplies & Engineering Solutions.
We record a functional test video before shipping.
✓ 3–12 Months Functional Warranty on New, Surplus & Refurbished Parts.
Returns: No-hassle return and replacement policy.
Quick Quotes: Receive Price & Availability Today! Email sales@globalvalvehub.com or WhatsApp +1 332 322 6825.
Payment Options: Wire Transfer (T/T) PayPal Visa MasterCard