Taxonomy Path: Analyzer Systems > Liquid Analytics > Liquid Analysis Sensors
Chlorine dioxide and ozone sensors in analog or digital form
The
JUMO tecLine ClO2 and O3
series (type 202634) groups eight membrane-covered amperometric sensors for chlorine dioxide and ozone in aqueous solutions. The chlorine dioxide sensors are intended for water from chlorite/chlorine and chlorite/hydrochloric acid generator plants; the ozone sensors can measure electrolytically generated ozone. Each sensor has integrated electronics that give a temperature-compensated 4 to 20 mA signal or a digital Modbus RTU signal. The standard versions use a hydrophobic PTFE membrane, whereas the chemical-resistant versions use an elastomer membrane that tolerates surfactants, bottle-washer and CIP chemicals and seawater. Like all JUMO membrane sensors in the family they cannot prove the absence of the measured disinfectant.
Technical specifications
Parameter
Specification
Type
202634 JUMO tecLine ClO2 + O3, two-electrode amperometric, membrane-covered
Versions
/45 ClO2 4-20 mA; /47 ClO2 4-20 mA chemical-resistant; /50 O3 4-20 mA; /52 O3 4-20 mA chemical-resistant; /60 O3 digital; /62 O3 digital chemical-resistant; /65 ClO2 digital; /67 ClO2 digital chemical-resistant
Ranges (order code)
10: 0 to 0.5; 20: 0 to 2; 25: 0 to 5; 35: 0 to 10; 37: 0 to 20 mg/l (ppm); /45 offers 0.5, 2, 5, 10 mg/l; /65 offers 2 and 20 mg/l
Resolution
0.001 mg/l on 0.5 or 2 mg/l ranges; 0.01 mg/l on higher ranges
Supply
Analog: DC 12 to 30 V, burden (UB – 11 V)/0.02 A, galvanic isolation required; digital: DC 9 to 30 V
Connection
Analog: 2-pole terminals 2 x 1 mm2; digital: 5-pole M12 flange connector, Modbus RTU
Response t90
About 15 s (standard ClO2 versions); about 60 s (chemical-resistant ClO2 version /47)
Settling and drift
About 1 h at first start; slope drift about -1 % per month (25 °C, pH 7.2)
Measuring water
0 to 45 °C standard (0 to 50 °C for /47); environment 0 to 55 °C; no ice crystals
pH range
pH 1 to pH 12; zero-point adjustment not required
Cross-sensitivity (ClO2 sensors)
Cl2 factor
| Delivery: | UPS, FedEx and DHL Express International delivery available Same day shipping for in-stock items on location Will negotiate custom freight & clearance details |
|---|---|
| Combined Shipping Policy: | We combine shipments to reduce costs for customers purchasing multiple items. After completing your selection, request a consolidated proforma invoice including shipping and handling. |
| Warranty: | All products are covered by Global Valve Hub — Industrial Supplies & Engineering Solutions. We record a functional test video before shipping. ✓ 3–12 Months Functional Warranty on New, Surplus & Refurbished Parts. |
| Returns: | No-hassle return and replacement policy. |
| Quick Quotes: | Receive Price & Availability Today! Email sales@globalvalvehub.com or WhatsApp +1 332 322 6825. |
| Payment Options: | Wire Transfer (T/T) PayPal Visa MasterCard |