LAR QuickCODultra Online COD (TOD) Analyzer


  • Part Number: LAR
  • Category: Analyzer Systems
  • Manufacturer: LAR Process Analysers
  • Product Origin: Germany / EU Germany / EU
  • Issue Date: 2026-11-26 17:35:17
  • SKU: IS-102475
  • ✓Certified Refurbished & Surplus — Save 30-60% vs New Price
  • ✓Fully Tested & Calibrated — 100% Functional Test & Report Included
  • ✓3-12 Months Warranty — Guaranteed Functional Warranty
  • ✓Fast Global Express Shipping — 3-7 Days Delivery via DHL / FedEx / UPS
Inquire via WhatsApp
New Product
On Request
✓ Original Factory Packaging
✓ Ship by: 2-4 Weeks
Never Used W/O Package
On Request
✓ Open-Box Surplus Stock
✓ Protective Industrial Packaging
✓ Ship by: 3-7 Business Days
Refurbish Grade A
On Request
✓ Inspected & 100% Function Tested
✓ 90%+ Like New Appearance
✓ Certified Refurbished
✓ Ship by: 5-10 Business Days
Refurbish Grade B
On Request
✓ Used, Repaired & Function Tested
✓ Ship by: 4-10 Business Days

Taxonomy Path: Analyzer Systems > Liquid Analytics > Liquid Analytical Systems

Overview
The
LAR QuickCODultra
is an online COD (TOD) water analyzer developed by LAR Process Analysers in Berlin, now part of Process Insights. It measures chemical oxygen demand through total oxygen demand, by oxidizing the complete water sample at 1,200 °C and detecting the oxygen consumed. The method follows ASTM D6238 and needs no catalysts and no additional hazardous chemicals, unlike the dichromate reflux method used in laboratories. A COD (TOD) result is available within 1 to 2 minutes, and the analyzer can handle sticky, fatty, salty and particle-laden samples, including salt concentrations up to 600 g/l.
our engineering team supplies the LAR QuickCODultra to industrial wastewater treatment plants, refineries, petrochemical complexes and municipal operators in Oman and the GCC. From Muscat we supply consumables on request and provide engineering support with application questionnaires, sample preparation and integration into plant control systems.
Measurement Principle
In the LAR QuickCODultra a robotic injection system doses a defined sample volume into a high-temperature reactor held at 1,200 °C, where all oxidizable organic and inorganic compounds are converted completely. A carrier gas with a known oxygen content flows through the reactor, and the analyzer measures the oxygen consumed by the oxidation, which is the total oxygen demand of the sample. Because oxidation is thermal rather than chemical, results are free of chloride interference thanks to LAR process management, and there is no reagent handling or chromium and mercury waste.
In general, TOD and COD correlate closely for most wastewater matrices, which is why a TOD-based analyzer is widely used for continuous COD monitoring and discharge control. The LAR QuickCODultra works on a batch principle, and optional sum parameters such as TOC and TNb can be measured in combination with COD (TOD), giving a fuller picture of organic and nitrogen loads from one instrument.
Engineering Design of the LAR QuickC

Delivery: UPS, FedEx and DHL Express
International delivery available
Same day shipping for in-stock items on location
Will negotiate custom freight & clearance details
Combined Shipping Policy: We combine shipments to reduce costs for customers purchasing multiple items. After completing your selection, request a consolidated proforma invoice including shipping and handling.
Warranty: All products are covered by Global Valve Hub — Industrial Supplies & Engineering Solutions.
We record a functional test video before shipping.
✓ 3–12 Months Functional Warranty on New, Surplus & Refurbished Parts.
Returns: No-hassle return and replacement policy.
Quick Quotes: Receive Price & Availability Today! Email sales@globalvalvehub.com or WhatsApp +1 332 322 6825.
Payment Options: Wire Transfer (T/T) PayPal Visa MasterCard

Related Products

LAR
Request Inquiry
More Info
LAR
Request Inquiry
More Info
LAR
Request Inquiry
More Info
LAR
Request Inquiry
More Info
LAR
Request Inquiry
More Info
LAR
Request Inquiry
More Info
↑