Taxonomy Path: Analyzer Systems > Gas Analytics > Gas Analyzer Systems
Michell XTC601-GP1 thermal conductivity binary gas analyzer, general purpose
The Michell XTC601-GP1 is the general-purpose (safe-area) wall-mounted version of the XTC601 binary gas analyzer. It determines the percentage of one gas in a binary or pseudo-binary mixture, such as hydrogen purity in air, CO2 in methane or helium recovery, by comparing the thermal conductivity of the sample against a sealed reference chamber. Matched thermistors in a Wheatstone bridge, inside a temperature-controlled isothermal cell, give a stable signal. our engineering team is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.
What XTC601-GP1 includes as standard
The code covers a backlit LCD with four capacitive buttons that work through the glass, an IP66 wall-mounted enclosure, a 316 stainless steel sample path, two 4 to 20 mA outputs, Modbus RTU over RS485, two concentration alarms, three M20 cable entries and 24 V DC power. The gas connection is 1/4 inch NPT female. Local screens show concentration with a status bar, a graph of the latest period, minimum and maximum values, readings from an external sensor and the alarm history. A light guide follows NAMUR 44.
XTC601-GP1 technical specifications
Part number
XTC601-GP1
Gas connections
1/4 inch NPT female
Sample flow (general purpose)
100 to 500 ml/min
Ambient temperature
+5 to +60 °C
Sample temperature
constant +5 to +45/+55 °C
Sample cell temperature
+50 or +60 °C
Warm-up / stabilisation
under 25 minutes at 20 °C / 5 minutes
Measured gases
air, argon, methane, CO2, hydrogen, helium or nitrogen (order codes add NMHC and CO)
Measurement range
selectable from 0 to 1 percent up to 0 to 100 percent, or 50 to 100 up to 98 to 100 percent
Accuracy
better than ±1 percent of range or ±0.05 percent H2 or He, whichever is greater
Response T90
under 50 seconds; under 20 seconds for H2 or He
Repeatability / linearity
±0.2 percent / ±1 percent of range
Zero and range stability
±0.5 percent of range per month e
| Delivery: | UPS, FedEx and DHL Express International delivery available Same day shipping for in-stock items on location Will negotiate custom freight & clearance details |
|---|---|
| Combined Shipping Policy: | We combine shipments to reduce costs for customers purchasing multiple items. After completing your selection, request a consolidated proforma invoice including shipping and handling. |
| Warranty: | All products are covered by Global Valve Hub — Industrial Supplies & Engineering Solutions. We record a functional test video before shipping. ✓ 3–12 Months Functional Warranty on New, Surplus & Refurbished Parts. |
| Returns: | No-hassle return and replacement policy. |
| Quick Quotes: | Receive Price & Availability Today! Email sales@globalvalvehub.com or WhatsApp +1 332 322 6825. |
| Payment Options: | Wire Transfer (T/T) PayPal Visa MasterCard |