Swan AMI Hydrazine Analyzer for Boiler Feedwater Oxygen Scavengers


  • Part Number: Swan
  • Category: Analyzer Systems
  • Manufacturer: Swan Analytical
  • Product Origin: Germany / EU Germany / EU
  • Issue Date: 2026-11-26 17:35:17
  • SKU: IS-104333
  • ✓Certified Refurbished & Surplus — Save 30-60% vs New Price
  • ✓Fully Tested & Calibrated — 100% Functional Test & Report Included
  • ✓3-12 Months Warranty — Guaranteed Functional Warranty
  • ✓Fast Global Express Shipping — 3-7 Days Delivery via DHL / FedEx / UPS
Inquire via WhatsApp
New Product
On Request
✓ Original Factory Packaging
✓ Ship by: 2-4 Weeks
Never Used W/O Package
On Request
✓ Open-Box Surplus Stock
✓ Protective Industrial Packaging
✓ Ship by: 3-7 Business Days
Refurbish Grade A
On Request
✓ Inspected & 100% Function Tested
✓ 90%+ Like New Appearance
✓ Certified Refurbished
✓ Ship by: 5-10 Business Days
Refurbish Grade B
On Request
✓ Used, Repaired & Function Tested
✓ Ship by: 4-10 Business Days

Taxonomy Path: Analyzer Systems > Liquid Analytics > Liquid Analytical Systems

Hydrazine and carbohydrazide monitoring in power plant feedwater
The
Swan AMI Hydrazine
is a microprocessor-controlled system for the determination and control of hydrazine or carbohydrazide used as boiler feedwater oxygen scavengers. In power plants hydrazine removes oxygen that could cause corrosion, and the reaction of hydrazine with oxygen produces nitrogen and water. The swan.ch page lists the product as AMI Hydrazin and describes a membrane-free, amperometric instrument.
The sensor uses three-electrode amperometry. A voltage is applied to two platinum electrodes and held at an optimum level by the reference electrode. Hydrazine in the sample generates a small current at the platinum pin electrode that is proportional to its concentration. A rotor cleans the electrode surfaces hydrodynamically and a Hall sensor on the rotor triggers a flow alarm if its speed drops.
Technical specifications
Values are from the Operator’s Manual A-96.250.501, section 2.2.
Parameter
Specification
Measuring range
0.1 to 600 ppb hydrazine
Accuracy
5% of reading up to 200 ppb; +/-15% up to 600 ppb, or +/-2 ppb, whichever is greater
Stability
+/-5% of reading per month, or +/-2 ppb per month
Response time
90% of a change within 60 seconds after the sample enters the flow cell
Temperature measurement
Up to 60 C, resolution 0.1 C
Sample flow
About 15 l/h
Sample temperature and pressure
15 to 45 C; inlet 0.15 to 2 bar; outlet pressure-free
Sample pH
Equal to or higher than 7.0
Reagent consumption
Less than 1 l of diisopropylamine per month at 25 C
Connections
Sample inlet Serto 4 x 6 mm; outlet tube adapter 15 x 20 mm
Power supply
100-240 VAC (+/-10%), 50/60 Hz, or 10-36 VDC; maximum 35 VA
Transmitter
Aluminum, IP 66 / NEMA 4X; -10 to +50 C; backlit LCD 75 x 45 mm
Panel
Stainless steel, 280 x 850 x 180 mm, 12.0 kg; mounting hole distance 254 x 824 mm
Fluidics and pH buffering
Hydrazine measurement requires a high pH. The AMI Hydrazine buffers the sample to above pH 10.5 by adding diisopr

Delivery: UPS, FedEx and DHL Express
International delivery available
Same day shipping for in-stock items on location
Will negotiate custom freight & clearance details
Combined Shipping Policy: We combine shipments to reduce costs for customers purchasing multiple items. After completing your selection, request a consolidated proforma invoice including shipping and handling.
Warranty: All products are covered by Global Valve Hub — Industrial Supplies & Engineering Solutions.
We record a functional test video before shipping.
✓ 3–12 Months Functional Warranty on New, Surplus & Refurbished Parts.
Returns: No-hassle return and replacement policy.
Quick Quotes: Receive Price & Availability Today! Email sales@globalvalvehub.com or WhatsApp +1 332 322 6825.
Payment Options: Wire Transfer (T/T) PayPal Visa MasterCard

Related Products

Monitor
Request Inquiry
More Info
Swan
Request Inquiry
More Info
Swan
Request Inquiry
More Info
Swan
Request Inquiry
More Info
Swan
Request Inquiry
More Info
Swan
Request Inquiry
More Info
↑