Taxonomy Path: Portable Test & Measurement > Process, Pressure & Temperature Calibrators
WIKA CTD9100 temperature dry-well calibrator, industrial series from -50 to +650 °C
The
WIKA CTD9100
series is a family of portable dry-well calibrators for checking and adjusting industrial temperature sensors on site or in a laboratory. This data sheet covers four models: CTD9100-COOL (-50 … +200 °C), CTD9100-165 (-35 … +165 °C), CTD9100-450 (+40 … +450 °C) and CTD9100-650 (+40 … +650 °C), together covering -50 to +650 °C [-58 to +1,202 °F]. Measurement uncertainties are quoted from 0.15 to 0.8 K. our engineering team supplies the CTD9100 in Oman and the GCC as an independent supplier in Muscat, and this page explains the four models so an RFQ can name the right one and its inserts.
How a dry-well calibrator like the CTD9100 works
A dry-well calibrator heats (and in some models cools) a temperature-controlled metal block with drilled inserts. A thermometer under test is inserted to the bottom of the bore and the block is brought to a set temperature, which is controlled by a built-in Pt100 reference sensor. When the block is stable, the reading of the thermometer under test is compared with the display of the calibrator, so deviations can be found or the thermometer adjusted. In the CTD9100, the actual and set temperature of the block appear at the same time on a large four-digit high-contrast LED display, and the set point is changed with two buttons. The immersion depth is 150 mm [5.91 in], which reduces heat conduction error along the sensor stem; a new block design improves temperature homogeneity at the lower range.
Technical specifications of the CTD9100 models
Parameter
COOL
165
450
650
Manufacturer / part number
WIKA / CTD9100-COOL
WIKA / CTD9100-165
WIKA / CTD9100-450
WIKA / CTD9100-650
Temperature range
-50 … +200 °C
-35 … +165 °C (-30 for X version)
+40 … +450 °C
+40 … +650 °C
Heating principle
Peltier element
Peltier element
Resistance heating
Resistance heating
Accuracy
±0.25 K
±0.20 K
±0.30 K
±0.50 K
Stability
±0.05 K
±0.05 K
±0.05 K to 100 °C; ±0.1 K to 4
| Delivery: | UPS, FedEx and DHL Express International delivery available Same day shipping for in-stock items on location Will negotiate custom freight & clearance details |
|---|---|
| Combined Shipping Policy: | We combine shipments to reduce costs for customers purchasing multiple items. After completing your selection, request a consolidated proforma invoice including shipping and handling. |
| Warranty: | All products are covered by Global Valve Hub — Industrial Supplies & Engineering Solutions. We record a functional test video before shipping. ✓ 3–12 Months Functional Warranty on New, Surplus & Refurbished Parts. |
| Returns: | No-hassle return and replacement policy. |
| Quick Quotes: | Receive Price & Availability Today! Email sales@globalvalvehub.com or WhatsApp +1 332 322 6825. |
| Payment Options: | Wire Transfer (T/T) PayPal Visa MasterCard |