Yokogawa PH450G-A-A pH/ORP Converter, AC


  • Part Number: PH450G-A-A
  • Category: Analyzer Systems
  • Manufacturer: Yokogawa
  • Product Origin: Germany / EU Germany / EU
  • Issue Date: 2026-11-26 17:35:33
  • SKU: IS-113090
  • ✓Certified Refurbished & Surplus — Save 30-60% vs New Price
  • ✓Fully Tested & Calibrated — 100% Functional Test & Report Included
  • ✓3-12 Months Warranty — Guaranteed Functional Warranty
  • ✓Fast Global Express Shipping — 3-7 Days Delivery via DHL / FedEx / UPS
Inquire via WhatsApp
New Product
On Request
✓ Original Factory Packaging
✓ Ship by: 2-4 Weeks
Never Used W/O Package
On Request
✓ Open-Box Surplus Stock
✓ Protective Industrial Packaging
✓ Ship by: 3-7 Business Days
Refurbish Grade A
On Request
✓ Inspected & 100% Function Tested
✓ 90%+ Like New Appearance
✓ Certified Refurbished
✓ Ship by: 5-10 Business Days
Refurbish Grade B
On Request
✓ Used, Repaired & Function Tested
✓ Ship by: 4-10 Business Days

Taxonomy Path: Analyzer Systems > Liquid Analytics > Liquid Analytical Systems

Introduction to the Yokogawa PH450G-A-A
The pH loop is the place where a converter earns its keep: glass impedance, long cables and wet terminal boxes all conspire against the signal. Yokogawa PH450G-A-A is the AC-powered general purpose pH and ORP converter of the EXAxt 450 line, designed with a dual high-impedance amplifier and a liquid earth so that the measuring electrode and the reference are treated symmetrically.
We list it because maintenance teams in Oman tend to look for the PH450G code by its exact string when a field unit fails; the suffix pair A-A identifies the mains version with cable glands.
our engineering team is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.
What PH450G-A-A can read
Parameter
Value
pH input range
-2 to 16 pH
ORP input
-1500 to 1500 mV
rH
0 to 100 rH
Input impedance
Dual high impedance, at least 10 to the 12 ohm
Temperature elements
Pt1000, Pt100, 350 ohm (DKK), 5k1, 6k8, PTC 10k, 3k Balco: -30 to 140 C; NTC 8k55: -10 to 120 C
Accuracy
pH within 0.01; ORP within 1 mV; temperature within 0.3 C
Response
Less than 4 s for 90 percent (pH 7 to pH 4)
Outputs, relays and wash
Two isolated 4-20 mA outputs, common negative, maximum load 600 ohm, with bi-directional HART 5 on mA1.
Output function linear or a 21-step table for pH, temperature, ORP or rH, with PID control and burn-up 21.0 mA or burn-down 3.6 mA per NAMUR NE43.
Four SPDT relays rated 100 VA at 250 V AC 5 A; they can signal high or low alarms, run on-off, duty-cycle or pulse-frequency control, trigger wash cycles, show hold, and S4 serves as the fail-safe contact.
Why the diagnostics matter
The converter performs sensor diagnostics by monitoring impedance, which means a cracked glass bulb or a blocked reference junction is flagged by the instrument, not discovered during a shift audit. Temperature compensation follows the Nernst equation with processing options: a configurable temperature coefficient, NEN6411 for water or strong acids and

Delivery: UPS, FedEx and DHL Express
International delivery available
Same day shipping for in-stock items on location
Will negotiate custom freight & clearance details
Combined Shipping Policy: We combine shipments to reduce costs for customers purchasing multiple items. After completing your selection, request a consolidated proforma invoice including shipping and handling.
Warranty: All products are covered by Global Valve Hub — Industrial Supplies & Engineering Solutions.
We record a functional test video before shipping.
✓ 3–12 Months Functional Warranty on New, Surplus & Refurbished Parts.
Returns: No-hassle return and replacement policy.
Quick Quotes: Receive Price & Availability Today! Email sales@globalvalvehub.com or WhatsApp +1 332 322 6825.
Payment Options: Wire Transfer (T/T) PayPal Visa MasterCard

Related Products

EJX115A
Request Inquiry
More Info
EJX118B
Request Inquiry
More Info
UT55A-000-10-00
Request Inquiry
More Info
EJX438A
Request Inquiry
More Info
PH8ERP-05-TN-N-E
Request Inquiry
More Info
↑