Yokogawa SC450G-D-A Conductivity Converter, 24 V DC


  • Part Number: SC450G-D-A
  • Category: Analyzer Systems
  • Manufacturer: Yokogawa
  • Product Origin: Germany / EU Germany / EU
  • Issue Date: 2026-11-26 17:35:33
  • SKU: IS-113088
  • ✓Certified Refurbished & Surplus — Save 30-60% vs New Price
  • ✓Fully Tested & Calibrated — 100% Functional Test & Report Included
  • ✓3-12 Months Warranty — Guaranteed Functional Warranty
  • ✓Fast Global Express Shipping — 3-7 Days Delivery via DHL / FedEx / UPS
Inquire via WhatsApp
New Product
On Request
✓ Original Factory Packaging
✓ Ship by: 2-4 Weeks
Never Used W/O Package
On Request
✓ Open-Box Surplus Stock
✓ Protective Industrial Packaging
✓ Ship by: 3-7 Business Days
Refurbish Grade A
On Request
✓ Inspected & 100% Function Tested
✓ 90%+ Like New Appearance
✓ Certified Refurbished
✓ Ship by: 5-10 Business Days
Refurbish Grade B
On Request
✓ Used, Repaired & Function Tested
✓ Ship by: 4-10 Business Days

Taxonomy Path: Analyzer Systems > Liquid Analytics > Liquid Analytical Systems

DC-powered Yokogawa SC450G-D-A at a glance
Some process sites cannot offer mains power at the analyser, and for them Yokogawa SC450G-D-A exists. It is the DC-supplied general purpose build of the EXAxt 450 four-wire conductivity converter: same touchscreen, same measurement engine, same relay and HART capability as the AC unit, but powered from 12 to 24 V DC.
Think of it as the answer to a question about the cabinet rather than the sensor. If the plant has a 24 V DC distribution with battery backup, or the converter sits on a packaged water-treatment skid, SC450G-D-A removes the need for a separate mains cable.
our engineering team is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.
What the SC450G-D-A suffix tells you
Parameter
Value
D
DC version: 12 to 24 V DC rating, 10.8 to 26.4 V DC acceptable, 10 W
A (second)
General purpose type with IP66 cable glands supplied
Not included
FM nonincendive approval, which belongs to the U type
Options possible
/UM, /U, /PM, /H5, /AFTG, /ANSI, /AM20, /SCT
Reading the code this way avoids ordering mistakes. A buyer who needs a 24 V DC unit and writes only SC450G on the enquiry could receive an AC unit, so the D matters.
Measurement capability
The input stage accepts two-electrode and four-electrode cells with cell constants from 0.005 up to 50.0 per cm and works with cable up to 60 m (200 ft) of WU40 or WF10. Conductivity is displayed from 0.000 uS/cm to 2000 mS/cm and resistivity up to 1000 Mohm-cm. Five temperature elements are supported for compensation: Pt1000 from -20 to 250 C, Pt100 and Ni100 from -20 to 200 C, NTC 8k55 from -10 to 120 C and Pb36 from -20 to 120 C. The reference temperature is programmable from 0 to 100 C, with 25 C as default.
Control, relays and signals
Output function can be linear or a 21-step non-linear table for conductivity, resistivity, concentration or temperature.
PID control may act on either the mA outputs or the contact outputs, and pulse frequency or pul

Delivery: UPS, FedEx and DHL Express
International delivery available
Same day shipping for in-stock items on location
Will negotiate custom freight & clearance details
Combined Shipping Policy: We combine shipments to reduce costs for customers purchasing multiple items. After completing your selection, request a consolidated proforma invoice including shipping and handling.
Warranty: All products are covered by Global Valve Hub — Industrial Supplies & Engineering Solutions.
We record a functional test video before shipping.
✓ 3–12 Months Functional Warranty on New, Surplus & Refurbished Parts.
Returns: No-hassle return and replacement policy.
Quick Quotes: Receive Price & Availability Today! Email sales@globalvalvehub.com or WhatsApp +1 332 322 6825.